
This figure shows a simplified block diagram of the apparatus. The diagram
does not show all the functions of each module, but it does represent the
most important functions of each modular component of the spectrometer.
The pulse programmer creates the pulse stream that gates the synthesized
oscillator into radio frequency pulse bursts, as well as triggering the
oscilloscope on the appropriate pulse. The rf pulse bursts are amplified
and sent to the transmitter in the sample probe. The rf current bursts
in these coil produce a homogeneous 12 gauss oscillating magnetic field
at the sample. These are the time dependent B fields that produce the precession
of the magnetization, referred to as the 90o (or p/2)
and 180o (or p) pulses. The transmitter
coils are wound in a Helmholz configuration to optimize rf magnetic field
homogeneity.
Nuclear magnetization precessing in the direction transverse to the applied constant magnetic field induces an EMF in the receiver coil, which is then amplified by the receiver circuitry. This amplified radio frequency (15 MHz) signal can be detected (demodulated) by two separate and different detectors. The rf amplitude detector rectifies the signal and has an output proportional to the peak amplitude of the rf precessional signal.
The other detector is a mixer, which effectively multiplies the precession signal from the sample magnetization with the master oscillator. Its output frequency is proportional to the difference between the two frequencies. This mixer is essential for determining the proper frequency of the NMR signal. The precession frequency of the nuclei is well defined by the local magnetic field and the nuclear magnetic moment. The oscillator is tuned to this precession frequency when a zero-beat output signal of the mixer is obtained. A dual channel scope allows simultaneous observations of the signals from both detectors,
The magnetic field is variable over a wide range. To detect the H and F signals, only lower currents (less than a third of the maximum) are needed.
The magnet and the power supply need water cooling. Before switching on the power supply, make sure that the water flows and the current setting is "0".
Fine tuning of the magnetic field is possible by adjusting the knob at the upper right corner. Notice that the fine tuning jumps at +50% to -50%. The field may exhibit a small drift over time. For studies when this may be an important issue, make sure that you run control tests from time to time, so that the drift is known and corrected for.
Two additional coils, connected to a separate (smaller) power supply, are installed to the magnet poles. The winding of these coils is opposite, and they can be used to create a magnetic field inhomogeneity.
The case for the modules has a fused and switched power entry unit located on the back right side. The unit uses 2 A slow blow fuses. A spare set of fuses is stored inside the fuse case. The spectrometer case has a linear power supply enclosed. It has slots for five modules, which connect to the power supply through a back plane of electrical connectors. The modules should be located as follows:
| Receiver | Pulse programmer |
Oscillator amplifier mixer |
Blank | Blank |
This
is a low noise, high gain, 15 MHz receiver designed to recover rapidly
from an overload and to amplify the radio frequency induced EMF from the
precessing magnetization. The input of the receiver is connected directly
to a high Q coil wrapped around the sample vials inside the sample probe.
The tiny induced voltage from the precessing spins is amplified and detected
inside this module. The module provides both the amplified rf signal as
well as detected signal.
The detector output rectifies the signal ("cuts" it in half) and passes only the envelope of the rf signal. It is also important to remember that the precession signal from spin system cannot be observed during the rf pulse from the oscillator/amplifier since these transmitter pulses induce voltages in the receiver coil on the order of 10 volts and the nuclear magnetization creates induced EMFs of about 10 mV; a factor of 106 smaller!
The pulse programmer
is a complete, self contained pulse generator which creates the pulse sequences
used in all the experiments. The pulses can be varied in width (pulse duration),
spacing, number, and repetition time. Pulses are about 4 V positive pulses
with a rise time of about 15 ns. The controls and connectors are described
below and pictured here
Accuracy: 1 part in 106 on all delay times.
There
are three separate functioning units inside this module. A tunable 15 MHz
oscillator, an rf power amplifier, and a mixer. The oscillator is digitally
synthesized and locked to a crystal oscillator so that its stability is
better then 1 part in 106 over 30 minutes. The frequency
in MHz is displayed on a seven digit LED readout at the top center of the
instrument. This radio frequency signal can be extracted as a continuous
signal (CW-RF out, switch on) or as rf pulse burst in to the transmitter
coil inside the sample probe.
The second unit is the power amplifier. It amplifies the pulse bursts to produce 12 Gauss radio frequency magnetic fields incident on the sample. It has a peak power output of about 150 watts.
The third unit is the mixer. It is a nonlinear device that effectively
multiplies the cw rf signal from the oscillator (w0)
with the rf signals from the precessing nuclear magnetization (w
= g B0).
If the two rf signals have different frequencies a beat structure will
be superimposed on the signal. The beat structure is clearly evident on
the lower trace of the signals from a free induction signal, shown here.
The
frequency of the beats equals to the absolute value of w
- w0 . When the frequency
of the precession signal is equal to the oscillator frequency, then the
beat frequency is zero, and the mixer output and the detector output from
the receiver module have identical shapes. At this setting the magnetic
field is tuned to frequency of the spectrometer.
Here is a list of the controls:
IMPORTANT: DO NOT OPERATE THE POWER AMPLIFIER WITHOUT ATTACHING TNC CABLE
FROM SAMPLE PROBE. DO NOT OPERATE THIS UNIT WITH PULSE DUTY CYCLES LARGER
THAN 1%. DUTY CYCLES OVER 1% WILL CAUSE OVERHEATING OF THE OUTPUT POWER
TRANSISTORS. SUCH OVERHEATING WILL AUTOMATICALLY SHUT DOWN THE AMPLIFIER
AND SET OFF A BUZZER ALARM. IT IS NECESSARY TO TURN OFF THE ENTIRE UNIT
TO RESET THE INSTRUMENT. POWER WILL AUTOMATICALLY BE SHUT OF TO THE AMPLIFIER
IN CASE OF OVERHEATING AND RESET ONLY AFTER THE INSTRUMENT HAS BEEN COMPLETELY
SHUT OFF AT THE AC POWER ENTRY.
The
figure shows an artist sketch of the sample probe. The transmitter coil
is wound in a Helmholz coil configuration so that the axis is perpendicular
to the constant magnetic field. The receiver pickup coil is wound in a
solenoid configuration tightly around the sample vial. The coil's axis
is also perpendicular to the magnetic field. The precessing magnetization
induces an EMF in this coil which is subsequently amplified by the circuitry
in the receiver. Both coaxial cables for the transmitter and receiver coils
are permanently mounted in the sample probe and should not be removed.
Caution should be exercised if the sample probe is opened since the wires
inside are delicate and easily damaged. Care should be exercised that no
foreign objects, especially magnetic objects are dropped inside the sample
probe. They can seriously degrade or damage the performance of the spectrometer.
The glass sample vial should be positioned so that the sample is at the center of the pickup coil. This can be achieved by adjusting the position of an O-ring at the top of the vial.
The oscilloscope has powerful and convenient numerical integration and fast Fourier transform (FFT) features. A short and sketchy introduction to the principles of the FFT is given in the "HP Product Note 54600-3: FFT Lab Experiments Notebook"
The oscilloscope is interfaced to a PC, and data from the scope can be transfered to the computer in two different formats: "screen shot" or numerical file. The corresponding menus in the HP 34810A BenchLink/Scope software are called IMAGE and WAVEFORM.
The screen shot is great for purposes of illustration. (This is how the spectra in this manual were made.) However, the data transferred this way ends up in a file foramt (*.PCX) that can not be processed numerically. The "wave" format allows exporting the data to an ASCII (or EXCEL) file, and it is recommended if further evaluation is planned.
Back to the main page